当前位置:上海境宣实业有限公司>>日本光明北川式气体检测管>>气体快速检测管>> 132SC1,3-二氯丙烷气体检测管132SC 1,3-Dichloropropane 1-10ppm
室内空气甲醛检测管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl气体测毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl气体检知管132SB VINYL CHLORID
1,3-二氯丙烷气体检测管132SC 1,3-Dichloropropane 1-10ppm
Tube 32SC
1. PERFORMANCE
1) Measuring range :0.4-12.0 ppm 0.2-6.0 ppm 0.1-3.0 ppm
Number of pump strokes 1(100ml) 2(200ml) 4(400ml)
2) Sampling time :3 minutes/2 pump strokes
3) Detectable limit :0.05 ppm(400ml)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(0~15℃)(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 2 pump strokes
8) Colour change :Greenish yellow→Pink
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Hydrogen chlorine is produced and PH indicator is discoloured.
CH2=CHCI+CrO3+H2SO4→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
PRETEAT TUBE DETECTOR TUBE
6.0
5.0
4.0
3.0
2.0
1.0
(White)
ppm
200ml
132SC VINYL CHLORIDE
Substance Interference ppm Coexistence
Acetylene 1% Lower readings are given.
Ethylene FIG.1 300 〃
Hydrogen chloride 200 Higher readings are given.
Chlorine Similar stain is produced. ch1lo0r×ideV cinoynlc. 〃
(NOTE)
In case of 1 or 4 pump strokes, following formula is available for the actual concentration.
Actual concentration=Temperature corrected value×2/Number of strokes
0
4.0
3.0
1.0
2.0
Ethylene〔ppm〕?
FIG.1 Influence of Ethylene